Gel Clot Lyophilized Amebocyte Lysate (LAL/TAL) Single Test in Ampoule
Bioendo manufactures Lyophilized Amebocyte Lysate (LAL/TAL) for endotoxin detection since 1978. Lyophilized Amebocyte Lysate (LAL/TAL) Single Test in Ampoule is the is the most economic way to do endotoxin detection. Bioendo’s Lyophilized Amebocyte Lysate (LAL/TAL) reagent features strong resistance to interference and strong improvement for heat stability.
1. Product information
Bioendo’s Lyophilized Amebocyte Lysate (LAL/TAL) reagent is the product manufactured with amebocyte from blood of Tachypleus tridentatus.
After rehydrating LAL/TAL reagent with Water for Becterial Endotoxin Test, sample could be added directly to the ampoule. Then incubate solution at 37ºC±1ºC for 60 ±2 minutes. Take each tube in turn directly from the incubator, and invert it through about 180º in one smooth motion. If a firm gel has formed that remains in place upon inversion, it means the sample contains endotoxin. None extra endotoxin-free tube is needed when do endotoxin detection with our Gel Clot Lyophilized Amebocyte Lysate (LAL/TAL) Single Test in Ampoule.
2. Product parameter
Sensitivity Range: 0.03, 0.06, 0.125, 0.25, 0.5, 1.0, 2.0 EU/ml
Single Test per Ampoule
Positive or Negative Results
Require Sample Volume: 0.1ml per Tes
3. Product feature and application
Single step endotoxin detection without expensive endotoxin assay instruments. Suitable for end-product endotoxin testing before product released. Low cost endotoxin assay for pharmaceutical factories and researchers.
4. Ordering information
Gel Clot Lyophilized Amebocyte Lysate (LAL/TAL) Single Test in Ampoule
Catalog Number | Sensitivity EU/ml | ml/ampoule | Test/ampoule | Ampoule/pack |
G010030 | 0.03 | 0.1 | Single Test Ampoule | 10 |
G010060 | 0.06 | |||
G010125 | 0.125 | |||
G010250 | 0.250 | |||
G010500 | 0.5 | |||
G050030 | 0.03 | 0.5 | 4 | |
G050060 | 0.06 | |||
G050125 | 0.125 | |||
G050250 | 0.250 | |||
G050500 | 0.5 |
Bioendo™ GS Endotoxin Assay Kit (Gel Clot Single Tests) 44 tests/Kit
Catalog Number | Sensitivity EU/ml | ml/ampoule | Test/ampoule | Ampoule/pack |
GS44010030 | 0.03 | 0.1 | Single Test Ampoule | 44 vials Gel Clot; 1 Vial CSE 10A; 5 Vials TRW 2 ml |
GS44010060 | 0.06 | |||
GS44010125 | 0.125 | |||
GS44010250 | 0.250 | |||
GS44010500 | 0.5 |
Storage Condition:
The LAL reagent is stable for more than two years at 2-8℃. Store LAL reagent at 2-8℃ , and keep LAL reagent in dark place. LAL Reagent Water is stable for more than three years at 2-20℃.
Note:
Lyophilized Amebocyte Lysate (LAL/TAL) reagent manufactured by Bioendo is made from amebocyte lysate from the horseshoe crab (Tachypleus tridentatus).
For inquiry please kindly email to : lisa@prolabios.com